ClearCanvas RIS/PACS v13.0 Team Edition Release Notes

Updated September 8, 2014

The following is a list of some known incompatibilities, defects and warnings that may affect use of the software. Where applicable, workarounds and guidance is provided. A note on versioning: because the various ClearCanvas projects share certain common libraries, version numbers are shared among various products such that common libraries have sequential versioning from release to release. Consequently, the products do not necessarily have sequential version numbers.  For the most up-to-date release notes for this version of the software, follow the link below.        

http://www.clearcanvas.ca/Portals/0/ClearCanvasFiles/Documentation/ReleaseNotes/13_0/Team

Change Log

Version

Changes

v.13

New Features/Improvements

  • MIP and MinIP in Workstation
  • Slab MPR in Workstation
  • Breast Tomosynthesis storage and viewing in Workstation
  • Enhanced CT and MR viewing in Workstation
  • Editing and appending of key images on Workstation
  • Reapplication of rules by ServerPartition in the ImageServer
  • Improved support for the ImageServer to server as a backup PACS including support for overwriting studies
  • Study level auto-routing rules

Notable Bug Fixes

  • 6696: DICOM Editor save button appears to not save modified values
  • 7818: Out Of Memory error when workstation attempts to send really large images
  • 8033: Edits made in DICOM Editor are not seen in viewer if study was previously viewed in Study Filters
  • 10424: All Shredhost services appear identical in Task Manager
  • 10914: Forward slashes in HP/layout name cause weird menu nesting in Workstation
  • 11343: Cannot abort closing Workstation if media writing in progress
  • 11374: Incorrect time since study time reported in Easel
  • 11526: Editing patient name in IS by changing caps only does not work
  • 11585: Adding a private tag column in Study Filters crashed the Workstation
  • 11588: Primary study not loaded for linked procedures when go to next item selected
  • 11589: RIS report printing should support appropriate letterhead for multi facility sites
  • 11593: Image with invalid DOB year will cause shredhost to fail
  • 11617: First click in Easel collapsed mode does not successfully select the series
  • 11777: Cannot delete studies in the ImageServer using the Web GUI
  • 11853: ImageServer install fails to create two stored procedures due to DB script case-sensitivity for column references
  • 12183: RISServer cannot recover from httplistener error
  • 12185: ImageServer slow in responding to queries compared to other PACS
  • 12298: RIS worklist search does not support names with accents
  • 12372: ImageServer date range validation removes the date entered by the user
  • 12373: ImageServer attempts compression on MR Spect non-images and fails
  • 12374: Calendar control in ImageServer study search page malfunctions after validation
  • 12399: Viewer freezes up when changing layout viewing a streamed study
  • 12404: Patient Name query against McKesson PACS doesn't work.  User should disable the AutoLeadingWildcard setting in the ClearCanvas.ImageViewer.Explorer.Dicom.DicomExplorerConfigurationSettings group.
  • 12488: RIS merge/unmerge orders can leave orders in a broken state.  Important: Unmerge functionality has been removed to prevent orders from entering a broken state.  The workaround is to cancel the order and place the order again.
  • 12494: XML Rule editor window has malfunctioning scroll bars in Chrome
  • 12512: History and FilesystemQueue tab does not record compression performed on study

v.10.1

New Features/Improvements

  • Hanging protocols will capture tiles
  • Overlays can be selected by display set and configured by modality
  • Reformat (In-place MPR)
  • Memory management, Portable Edition updates

Notable Bug Fixes

  • 6696: DICOM Editor save button appears to not save modified values
  • 8033: Edits made in DICOM Editor are not seen in Viewer if the study was previously viewed in Study Filters
  • 8266:  Large multi-frame images can now be written to media
  • 9336: Workstation no longer crashes when server and client clocks are out of sync
  • 10269: Memory management issues in cine and stacking
  • 10768: Portable Edition media does not auto-start and shows error about .NET 4
  • 10941: DICOM Print does not work with comma as the decimal symbol
  • 10943: Hanging protocols won't work if used and saved with different decimal symbols
  • 10946,10947: ImageServer rules processing may depend on server's decimal and digit grouping symbols
  • 10995: Issues related to the use of comma as the decimal mark
  • 11035: Invalid DICOM file generated for PDF attachments

 

Other

  • “Administrative Services" has been renamed "EnterpriseServer" to match the component name

v.9

New Features/Improvements

  • RIS
  • CE Mark

Notable Bug Fixes

  • 8940: JPEG Lossless Process 14 codecs corrupt pixel data when Bits Stored=8, Bits Allocated=16
  • 9594: Web GUI query very slow with many partitions

v.6.2

Notable Bug Fixes

  • 6682: Viewer cannot render monochrome images with "rogue" palette color LUT tags
  • 9509: Can't view PDF attachments from Portable Workstation
  • 9967: Images with impossible high bit cannot be rendered
  • 9968: Viewer can get into infinite retry loop when steaming images with LUT encoding errors
  • 9998: Viewer should not send Instance Availability in C-FIND request
  • 10001: Cannot distinguish between same day studies without study time
  • 10013: Limited Windows user account cannot open images due to AppData folder permission restriction

v.6.1

New Features/Improvements

  • International edition support
  • Spanish localization
  • Workstation rebranding support

Notable Bug Fixes

  • 9600: Study status is not displayed for non-admin users

v.6

New Features/Improvements

  • Fine Grained Access to IS partitions
  • Fine Grained Access in IS GUI
  • SaaS rebranding support
  • Web Service for Patient Updates and Merges

v.5

New Features/Improvements

  • WebPortal
  • Fine-grained access control
  • PDF document import
  • Spine labelling
  • Digital subtraction imaging
  • Standardized uptake value (SUV)
  • Diagnostic information package
  • Probe tool has changed - it always shows post-modality LUT values with units

Notable Bug Fixes

  • 4079: Viewer process is limited to 2GB of process memory even when /3GB switch is set
  • 4999: No priors displayed when there are multiple default servers and one fails to respond in time
  • 7910: MPR and Fusion do not account for varied Modality LUTs in source images
  • 9246: Small precision loss can occur when measuring area on large images
  • 9345: Remove AJAX extensions installer dependency

v.4

New Features/Improvements

  • Hanging protocols with support for related studies (priors)
  • Easel, easy image selection
  • 9052: Updated the application to address DICOM CP-586 to support the use of Pixel Spacing (0028,0030) for projection modalities. For projection modalities, the text overlay will display the type of calibration used.

Notable Bug Fixes

  • 1466: Crash on startup in certain versions of Windows XP (e.g. Chinese)
  • 8391: Key images fail to be created when empty Procedure Code Sequence is present in source image
  • 8447: Images with empty View Code Sequence cannot be opened

Known Incompatibilities

ID

Summary

Description

Component

792

Images don't display on grayscale monitors

The application does not support dynamic/static grayscale palettes on older Dome/Planar monitors/video cards. Images may appear blank or there may be rendering artifacts.  Thoroughly evaluate image quality with your display configuration to ensure that it is supported before deploying for clinical use.  Otherwise, newer display device(s) may be required.

Workstation

2920

Some controls appear black on some, usually older, Dome monitors

The application does not support all Dome monitors, particularly older models.  Thoroughly evaluate image quality with your display configuration to ensure that it is supported before deploying for clinical use.  Otherwise, newer display device(s) may be required.

Workstation

9308

Right-clicks do not work properly for the Webstation on the Mac Safari browser

Users attempting to use the WebPortal in Mac Safari will find certain "right-click" tools and features inaccessible and unusable. The WebPortal and Webviewer have only been tested with Internet Explorer. Install and use Internet Explorer.

WebPortal

Configuration Warnings

ID

Summary

Description

Component

7173

Configure settings does not list local settings in Enterprise Mode

For the Team Edition of the Workstation, settings that are stored in the Local configuration file are not displayed in the Configure Settings component. With the Personal Edition, these settings are displayed.  The settings must be manually edited from the application configuration file found in the installation directory.

Workstation

8121

Exception generated on running workstation from network

The workstation should be installed only on the local machine.

Workstation

8449

DICOM related settings are not migrated on upgrade

Some settings are not being migrated as of yet on upgrade. These settings include those in the sections ClearCanvas.Dicom.Network.NetworkSettings, ClearCanvas.Dicom.ServiceModel.Streaming.StreamingSettings, and ClearCanvas.Dicom.DicomSettings.

Workstation

Known Bugs

ID

Summary

Description

Component

3914

Deleting log file while application is running causes strange behaviour

Only workaround is to stop the Workstation Shred Host Service, close the Workstation application, delete the log, then restart the service.

Workstation

4458

Streamed images displayed incorrectly when the server is compressing the images with JPEG Lossless

During compression of a study, the study may show up incorrectly in a Workstation client.

ImageServer

4881

Workstation considers a pixel to be in an ROI if the top-left corner is contained in the region.

Pixels are only considered to be in a region of interest if the top left corner is included in the region drawn - i.e. pixels on the top and left edge of the region may be unexpectedly excluded and pixels on the bottom and right edges may be unexpectedly included. If precision down to the pixel is required for mean, standard deviation and/or SUV, draw the ROI to include or exclude the top left corner of the desired pixels. The probe tool can be used to determine where the top-left corner of a pixel is.

Workstation

5750

MPR fails to render 8 bit volumes

No workaround, but such data sets are very rare.

Workstation

6590

Several pages are not displayed properly if the AE title contains spaces

The Device and WorkQueue pages do not display properly when there is a space in the AE Title.

ImageServer

6785

Opening study after display configuration change shows black workspace

If this problem occurs after changing monitor configuration, restart the application.

Workstation

7765

Study Integrity Queue Searching not consistent

Patient ID and Accession Number searches in the study Integrity Queue only search against the original study.  They do not search against the Conflicting Study.

ImageServer

7862

Does not render overlays from compressed images if overlay is embedded in pixel data

Images like this are not DICOM compliant.  Configure the modality to use the overlay repeating groups instead of embedding overlays in pixel data.

Workstation

8133

Reference lines are not captured in clipboard images

No workaround.

Workstation

8188

ImageServer UI can time out immediately after user logs in when client and server system clocks don't match

If the client system's clock is further ahead than the timeout specified, the GUI will immediately timeout after the user logs into the system.  

It is recommended that a time synchronization service, such as NTP be configured on the server and client.

ImageServer

8226

Media writer provides no progress during "Creating disc image" stage, which can take a very long time

This stage may take a long time, depending on the size of the images being written. Wait.

Workstation

8262

ImageServer Web GUI does not immediately show modified message after restarting a modified ImageServer shredhost

The anti-tampering mechanism in the ImageServer does not immediately display to the user when the installation has been modified.  It requires either the server to be rebooted or IIS to be restarted before the ImageServer shows that the installation has been modified.

ImageServer

8394

Key Image display sets with multiple study references always show a placeholder image for images from the "other" study

When you capture key images from more than one study in a session, all captured key images are added to a new KO series in all of the affected studies. It is arguable whether or not this is the correct way to do it, but the reason for doing it was to alert the user that key images were created in the same session for multiple studies regardless of which study was (later) opened.

However, when you open one or all of the affected studies (even via multi-selection), only key images from the study to which the key image display set belongs are shown; others are placeholder images.

Workstation

8458

Remnant orphaned SQL Database files cause the installer to fail

Sometimes when SQL Server goes to delete a database it leaves behind the underlying database files, e.g. mdf, ldf, and ndf files, even though the database had been deleted.  When this occurs, there are problems installing the ImageServer properly.  The workaround is to delete the database files through the SQL Management Studio

ImageServer

8819

Viewer will not print/export Fusion images

There is an issue with exporting and printing Fusion images - these operations may or may not complete successfully.

To workaround, toggle the colour bar overlay off before attempting to print or export. Alternatively, consider taking a screen capture.

Workstation

9252

Workstation does not properly support pixel padding value

No workaround. Images with pixel padding values may not display as the manufacturer of the modality intended.

Workstation

9449

The Easy Image Selection tool (Easel) can push the last study out of view when there are many related studies.

In Easel, if there are many related studies, it is possible that hovering over certain studies will "push" the last study out of view. When examining a list of studies in Easel, the user should use the scrollbar as a cue to whether there are more studies in the list.

Workstation

9660

Auto-refresh turns itself off on the Work Queue page

The Auto-refresh feature may turn itself off. Until this defect is addressed, the user can reset the auto-refresh or refresh manually.

ImageServer

9850

Unable to log in if timeout is less than SessionTokenCachingTimeToLiveSeconds

SessionTokenCachingTimeToLiveSeconds, by default 45 seconds, must remain less than the User Session timeout value, by default 15 minutes. If this is not the case, it users will be locked out of the system. The only way to recover is to modify the database directly and change the value back - which we do not recommend you do without contacting support

EnterpriseServer

9852

Searching for accented characters requires an exact match

For example searching for "López" will not find "Lopez" and vice versa. This is not technically a defect, but the likelihood of inconsistent data is high and it makes searching more difficult. The workaround is to use the ? wildcard when searching for these potentially different characters.

Workstation

9975

"My Computer" can perform poorly when accessing folder containing many zip files

No workaround.

Workstation

10703

WebPortal does not check if the browser has a PDF viewer plugin

If a browser does not have a PDF viewer plugin, the user is prompted by the browser to save or open the file ("Report.pdf"). If the user chooses to save the file, which may contain confidential patient information, it will remain on the computer until it is explicitly deleted.

Saved files should be managed as one would manage other PHI. Installing a PDF reader like Adobe Acrobat, will allow users to view PDF attachments and manage the deletion of the file after viewing.

WebPortal

11023

 "Unable to connect to the server. Refresh the page when your connection is restored." is misleading

When the MaxConnections for the Webstation has been exceeded (100 by default), the user gets this misleading message when the network connection is apparently ok and other webstations are still running properly.

It is possible to adjust the number of MaxConnections in the shredhost config.

Webstation

11333

Workstation GUI appears incorrectly when Windows is configured to use 125%

The workaround is to change the setting back to 100% (96 DPI)

Workstation

11375

Starting a procedure before its scheduled time causes scheduling conflicts for new procedures.

If a future scheduled procedure is checked-in, the scheduled start date/time is updated with the current timestamp but the scheduled end time is not changed. When a user attempts to create a new procedure that is within that extended time span between the new start time and the original end time, the modality appears "occupied", causing the conflict dialog to appear. This will not be obvious from the calendar.

The workaround is to reschedule future procedures before starting them.

RIS

11634

DICOM Editor is unusable for sufficiently large multiframe images

Workstation sometimes becomes unresponsive and crashes when opening a large multiframe image in the DICOM Editor.

Workstation

11639

Cannot copy a subset of frames in Enhanced MR/CT studies to the Clipboard

On an Enhanced CR or MR, Copy Range of Images to Clipboard does not allow user to select a range of images (frames)

Workaround: If copying individual frames is essential, consider having the modality resend the study using a non-Enhanced SOP class

Workstation

11658

Reference Lines on Enhanced CT/MR always show frame #1

Reference lines always say "1" (image number) on Enhanced CT/MR

Workstation, Webstation DX

11680

PDF reports not viewable on Surface IE

On Microsoft Surface’s Internet Explorer, PDF reports are not viewable

Webstation, Webstation DX

11686

Some tools in the Webstation may become unresponsive when user switches to another application using ALT-Tab

Similar to 11688.  Webstation thinks that a shortcut is being activated.

Workaround: Press ALT again, or use the mouse to switch applications instead to avoid the problem entirely.

Webstation, Webstation DX

11688

Using keyboard to change active browser window causes Webstation to think a shortcut is being used

Workaround: Press CTRL again, or use the mouse to switch applications instead to avoid the problem entirely.

Webstation, Webstation DX

11695

Reformatting a fusion display set results in a reformat of the underlying CT only

Reformatting of fused series are not currently supported.

Webstation, Webstation DX

11707

Incorrect study status shown on the GUI when server rules are being re-applied to the study

While server rules are being re-applied to a study (by running the Partition Re-apply Rules service), the study status on the Studies page changes to "Online-Idle" with "Idle" being a link which user can click on. Clicking on the link will open the Work Queue page but nothing is shown on the screen because there's no related WQI for the study.

ImageServer

11732

Opening breast tomosynthesis images on the Portable Workstation is very slow and sluggish

Because of the size of the images, loading them from a CD is very slow.  Load time can be as much as 20 seconds per image.

Workstation, Portable Edition

11772

Webportal may crash if browser window is resized

If user resizes the browser window when images have been loaded, Silverlight may crash and user has to reload the WebPortal.  Only happens on Chrome.

Workarounds: Resize before opening studies, maximizing the window, or use a different browser

WebPortal

12555

ImageServer does not work properly when partition AE Title contains space characters

If the Partition AE Title contains a space character--which is allowed in DICOM--  streaming and the web viewer n the ImageServer will stop working.

Workaround: Do not use spaces in the Partition AE Title

ImageServer

12562

Cannot get Explorer back if you cancel out of closing Workstation

If you have unviewed layouts, close workstation but cancel out of the closing operation, Explorer will have been closed and you cannot get Explorer back.

Workstation